High frequency ICP® pressure sensor, 500 psi, 10 mV/psi, 3/8-24 mtg thd, accel. comp., ground isolated
- Measurement Range: 500 psi
(3450 kPa)
- Sensitivity: (±10%)10 mV/psi
(1.45 mV/kPa)
- Low Frequency Response: (-5%)0.01 Hz
- Resonant Frequency: >=500 kHz
(>=500 kHz)
- Electrical Connector: 10-32 Coaxial Jack
- Weight: 0.41 oz
(11.6 gm)
|
English: |
SI: |
|
Performance |
Measurement Range (for ±5V output) |
500 psi |
3450 kPa |
|
Useful Overrange (for ± 10V output) |
1000 psi |
6895 kPa |
[3]
|
Sensitivity (±10 %) |
10 mV/psi |
1.45 mV/kPa |
|
Maximum Pressure (static) |
10000 psi |
68950 kPa |
|
Resolution |
2 mpsi |
0.014 kPa |
[2]
|
Resonant Frequency |
≥500 kHz |
≥500 kHz |
|
Rise Time (Reflected) |
≤1.0 µ sec |
≤1.0 µ sec |
|
Low Frequency Response (-5 %) |
0.01 Hz |
0.01 Hz |
|
Non-Linearity |
≤1.0 % FS |
≤1.0 % FS |
[1]
|
Environmental |
Acceleration Sensitivity |
≤0.002 psi/g |
≤0.0014 kPa/(m/s²) |
|
Temperature Range (Operating) |
-100 to +275 °F |
-73 to +135 °C |
|
Temperature Coefficient of Sensitivity |
≤0.03 %/°F |
≤0.054 %/°C |
[2]
|
Maximum Flash Temperature |
3000 °F |
1650 °C |
|
Maximum Shock |
20000 g pk |
196000 m/s² pk |
|
Electrical |
Output Polarity (Positive Pressure) |
Positive |
Positive |
|
Discharge Time Constant (at room temp) |
≥50 sec |
≥50 sec |
|
Excitation Voltage |
20 to 30 VDC |
20 to 30 VDC |
|
Constant Current Excitation |
2 to 20 mA |
2 to 20 mA |
|
Output Impedance |
<100 Ohm |
<100 Ohm |
|
Output Bias Voltage |
8 to 14 VDC |
8 to 14 VDC |
|
Electrical Isolation |
100000000 Ohm |
100000000 Ohm |
|
Physical |
Sensing Geometry |
Compression |
Compression |
|
Sensing Element |
Quartz |
Quartz |
|
Housing Material |
Stainless Steel |
Stainless Steel |
|
Diaphragm |
Invar |
Invar |
|
Sealing |
Welded Hermetic |
Welded Hermetic |
|
Electrical Connector |
10-32 Coaxial Jack |
10-32 Coaxial Jack |
|
Weight |
0.41 oz |
11.6 gm |
|
* All specifications are at room temperature unless otherwise specified
.

Specification Notes
[1] Zero-based, least-squares, straight line method.
[2] Typical.
[3] For +10 volt output, minimum 24 VDC supply voltage required. Negative 10 volt output may be limited by output bias.
[2] Typical.
[4] See PCB Declaration of Conformance PS023 for details.
Accessories
|
|
|
Supplied |
065A03 |
Seal ring, 0.435" OD x 0.377" ID x 0.030" thk, brass (for Series 101 and 102, and Model 061A01) |
3 |
Optional Versions
( M ) Metric Mount
Accessories: |
Supplied |
065A40 |
Seal ring 0.435" OD x 0.397" ID x 0.030" thk brass |
3 |
( S ) Stainless Steel Diaphragm