Model : 1102-03A/LCS-2A | Strain Gage Load Cell

Click for lead times
canister load cell, 200 lbf (890 n)  rated capacity, 50% static overload protection, 2mv/v output, 1/4-28 unf-2b threads, pt02e-10-6p connector, aluminum construction
Add to Cart for Quote

Canister load cell, 200 lbf (890 N) rated capacity, 50% static overload protection, 2mV/V output, 1/4-28 UNF-2B threads, PT02E-10-6P connector, aluminum construction

  • Measurement Range: 200 lb ( 890 N)
  • Size - Diameter: 2.75 in ( 69.9 mm)
  • Size - Height: 1.5 in ( 38.1 mm)
  • Electrical Connector: PT02E-10-6P
English: SI:
Performance
Measurement Range 200 lb 890 N [ 1]
Sensitivity (±10 %) 2 mV/V 2 mV/V [ 2]
Non-Linearity ≤0.05 % FS ≤0.05 % FS
Hysteresis ≤0.05 % FS ≤0.05 % FS
Non-Repeatability ≤0.02 % FS ≤0.02 % FS
Creep (in 20 minutes) ≤.025 % ≤.025 %
Static Error Band ≤.05 % FS ≤.05 % FS
Eccentric Load Sensitivity ≤.25 %/in ≤.25 %/25.4mm
Resonant Frequency 5.4 kHz 5.4 kHz
Environmental
Overload Limit 300 lb 1334 N
Load Limit (Side Force, FX or FY) 200 lb 890 N [ 3]
Load Limit (Bending Moment, MX or MY) 200 in-lb 23 Nm [ 3]
Load Limit (Axial Torque, MZ) 65 in-lb 7 Nm [ 3]
Temperature Range (Operating) -65 to +200 °F -54 to +93 °C
Temperature Range (Compensated) +70 to +170 °F +21 to +77 °C
Temperature Effect on Output (Maximum) ±0.0015 %Reading/°F ±0.0027 %Reading/°C [ 4]
Temperature Effect on Zero Balance (Maximum) ±0.0008 %FS/°F ±0.0015 %FS/°C [ 4]
Electrical
Bridge Resistance 700 Ohm 700 Ohm [ 5]
Excitation Voltage (Recommended) 10 VDC 10 VDC [ 6]
Insulation Resistance >5000000000 Ohm >5000000000 Ohm
Zero Balance ≤1 % FS ≤1 % FS
Output Polarity +Tension +Tension
Physical
Size - Height 1.5 in 38.1 mm [ 7]
Size - Diameter 2.75 in 69.9 mm [ 7]
Weight 0.67 lb 304 g
Mounting Thread 1/4-28 Female No Metric Equivalent
Housing Material Painted Aluminum Painted Aluminum
Sensing Element Strain Gage Strain Gage
Deflection at Full Scale Capacity 0.003 in 0.08 mm
Electrical Connector PT02E-10-6P PT02E-10-6P
* All specifications are at room temperature unless otherwise specified
.




Specification Notes


[ 1] FS - Full Scale.
[ 2] RO - Rated Output.
[ 3] Singularly applied, i.e. no other extraneous loads.
[ 4] Over compensated operating temperature range.
[ 5] Nominal.
[ 6] Calibrated at 10 VDC, usable 5 to 20 VDC or VAC RMS.
[ 7] See Outline Drawing 51277 for Complete Dimensions


Accessories


Optional
181-012A Mating conn., 6-socket (F) w/strain relief for sensors with PT conn. 1
8311-01-10A Cable assembly for sensors with PT conn., 10-ft, PT conn. to pigtails, 4-cond. cable 1