vibration calibration standard system, 100 mv/g, 10-32 sensor mtg (ce approved)

calibration system

Model 394A11

  • IP

Vibration calibration standard system, 100 mV/g, 10-32 sensor mtg (CE approved)

    Specification Notes


    [1] Zero-based, least-squares, straight line method.
    [2] Typical.
    [3] All sensor specifications are valid only when used with matched power supply.
    [2] Typical.
    [4] Provided by 488A04.

    English: SI:
    Performance
    Sensitivity (±0.5 %) (Voltage) 100 mV/g 10.2 mV/(m/s²) [3]
    Measurement Range (±5V ) (Output) 50 ±g pk 490 ±m/s² pk
    Frequency Range (±2 %) 5 to 5000 Hz 5 to 5000 Hz
    Frequency Range (±5 %) 0.5 to 10000 Hz 0.5 to 10000 Hz
    Frequency Range (±10 %) 0.3 to 14000 Hz 0.3 to 14000 Hz
    Resonant Frequency (Mounted) ≥35 kHz ≥35 kHz
    Broadband Resolution (1 Hz) 0.004 g rms 0.039 m/s² rms [2]
    Amplitude Linearity ≤1 % ≤1 % [1]
    Transverse Sensitivity ≤3 % ≤3 %
    Sensitivity (Base Strain) 0.0001 g/µε 0.001 (m/s²)/µε
    Environmental
    Warm Up <12.0 sec <12.0 sec
    Temperature Range (Operating) -65 to +250 °F -54 to +121 °C
    Electrical
    Spectral Noise (1 Hz) 235 (µm/s²)/√Hz 2303 [2]
    Spectral Noise (10 Hz) 55 (µm/s²)/√Hz 539 [2]
    Spectral Noise (100 Hz) 18 (µm/s²)/√Hz 176 [2]
    Spectral Noise (1 kHz) 12 (µm/s²)/√Hz 118 [2]
    Physical
    Sensing Element Quartz/Shear Quartz/Shear
    Housing Material Stainless Steel/Welded Stainless Steel/Welded
    Size - Hex 1.19 in 30.2 mm
    Size - Height 1.5 in 38.1 mm
    Weight 6.2 oz 176 gm [2]
    Electrical Connector 10-32 Coaxial/Side 10-32 Coaxial/Side
    Mounting Thread (Standard) 1/4-28 UNF 1/4-28 UNF
    Mounting Thread (Unit Under Test) 10-32 UNF 10-32 UNF
    Mounting Torque 20 to 30 in-lb 225 to 338 N-cm
    Power (±5% ) (Required) 36 VDC 36 VDC [4]
    Size - Height 6.3 in 16 cm
    Size - Width 2.4 in 6.1 cm
    Size - Depth 11.0 in 28 cm
    Weight 1.51 lb 685 gm [2]
    Cable Connector (Input) BNC Jack/Rear BNC Jack/Rear
    Cable Connector (Output) BNC Jack/Rear BNC Jack/Rear

    * All specifications are at room temperature unless otherwise specified.

    Accessories


    Supplied
    ACS-1 Single axis amplitude response calibration from 10 Hz to upper 5% frequency, NIST traceable 1
    ACS-4 Single axis, low frequency phase and amplitude response cal from 0.5 to 10 Hz, (requires ACS-1 and minimum sensitivity of 20 mV/g) 1
    003C10 Low-noise coaxial cable, blue TFE jacket, 10-ft, 10-32 coaxial plug to BNC plug 1
    003D03 Low-noise coaxial cable, blue TFE jacket, 3-ft, BNC plug to BNC plug 1
    017AXX 1
    080A149 Calibration adaptor, 10-32 external to 5-40 internal thread, tungsten (for Model 394A11 calibration standard) 1
    301A11 Calibration ICP® accel., 100 mV/g, 50g, 0.5 to 10 kHz, 10-32 side conn., 1/4-28 thd on bottom, 10-32 thd on top 1
    482A23 1-channel, line-powered, ICP® sensor signal cond., BNC input/output conn. X0.9 to x1.1 gain, (for use in 394A10/ 394A11 systems) 1
    488B04 Universal AC power adapter, 100-240 VAC/ 50-60 Hz input, 33 VDC output, 5-pin DIN (for models 482C05 and 482C15), line cord 1
    ACS-10/ACS-4 1
    M080A149 Calibration adaptor, 10-32 external to M3x0.5 internal thread, tungsten (for Model 394A11 calibration standard) 1
    081A08 Mounting stud, 10-32 to 1/4-28 x 0.30" long, BeCu 2
    081A90 Mounting stud, 10-32 to 5-40 2
    081B05 Mounting stud, 10-32 to 10-32, 0.275" long, BeCu w/shoulder 2
    081B20 Mounting stud, 1/4-28 to 1/4-28 x 0.365" long, BeCu w/shoulder 2
    M081B05 Mounting stud, 10-32 to M6x0.75-6g, 0.275" long, BeCu 2
    M081B20 Metric mounting stud, 1/4-28 to M6 x 0.75, BeCu w/shoulder 2

    Optional Versions


    (M) Metric Mount

    English SI
    Sensitivity (&#177;0.5 ) 10 mV/(m/s²) 98 mV/g