| Performance |
| Maximum Load |
28.2 oz |
800 gm |
[1]
|
| Frequency Range (operating, 100 gram payload) |
5 to 10000 Hz |
300 to 600000 cpm |
|
| Accuracy (10 Hz to 10 khz) |
±3 % |
±3 % |
|
| Accuracy (5 Hz to 10 Hz) |
±5 % |
±5 % |
|
| Accuracy (Velocity (10 Hz to 1000 Hz)) |
±3 % |
±3 % |
|
| Accuracy (Displacement (30 Hz to 150 Hz)) |
±3 % |
±3 % |
|
| Amplitude Linearity (100 gram payload, 100 Hz) |
1 % |
1 % |
[2]
|
| Distortion (100 gram payload, 30 Hz to 2 kHz) |
<5 % |
<5 % |
[3]
|
| Maximum Amplitude (Acc 50 Hz,10 gram payload) |
20 g pk |
196 m/s^2 pk |
|
| Maximum Amplitude (Vel 50 Hz, 10 gram payload) |
20 in/sec pk |
500 mm/s pk |
|
| Maximum Amplitude (Displ 50 Hz, 10 gram payload) |
150 mils pk-pk |
3.8 mm pk-pk |
|
| Maximum Amplitude (Acc 50 Hz, 500 gram payload) |
2.5 g pk |
24.5 m/s² pk |
|
| Maximum Amplitude (Vel 50 Hz, 500 gram payload) |
3.5 in/s pk |
90 mm/s pk |
|
| Control Interface |
| Test Operation |
Manual (Closed Loop) or Semi-Automatic |
Manual (Closed Loop) or Semi-Automatic |
|
| Test Sensor In |
ICP, Voltage, Modulated Current, Chg, Diff Chg, Single-ended Chg (Ext Chg AMP Req'd |
ICP, Voltage, Modulated Current, Chg, Diff Chg, Single-ended Chg (Ext Chg AMP Req'd |
[5]
|
| Bias Fault Indication (ICP® Sensors) |
Yes |
Yes |
|
| External Source In (Max) |
1V AC RMS |
1V AC RMS |
|
| Monitor Reference Out |
10 mV/g (nominal), Quartz Reference Accelerometer, BNC Jack Output |
10 mV/g (nominal), Quartz Reference Accelerometer, BNC Jack Output |
|
| Display Units (Acceleration) |
g pk & RMS |
m/s² pk & RMS |
|
| Display Units (Velocity) |
in/sec pk & RMS |
mm/sec pk & RMS |
|
| Display Units (Displacement) |
mils pk-pk |
um pk-pk |
|
| Display Units (Frequency) |
Hz |
CPM |
|
| Test Sensor Sensitivity |
mV/EU, mA/EU, µA/EU, pC/EU |
mV/EU, mA/EU, µA/EU, pC/EU |
[4]
|
| Environmental |
| Temperature Range (Operating) |
32 to 122 °F |
0 to 50 °C |
|
| Electrical |
| Power Required (Internal Battery (sealed solid gel lead acid)) |
12 VDC, 4 amp hours |
12 VDC, 4 amp hours |
|
| Power Required (Input from Charger) |
18 VDC, 1A |
18 VDC, 1A |
|
| Power Required (AC Power (for recharging battery)) |
110-240 V, 50-60 Hz |
110-240 V, 50-60 Hz |
|
| Battery Life (100 gram payload, 100 Hz 1g pk) |
18 hours |
18 hours |
[6]
|
| Battery Life (100 gram payload, 100 Hz 10g pk) |
1 hour |
1 hour |
[6]
|
| Physical |
| Storage |
Semi-Automated test routine calibration settings |
Semi-Automated test routine calibration settings |
|
| Points Per Record |
30 calibration data points |
30 calibration data points |
|
| Sensor Information |
Model number, serial number, Mounting orientation (x, y, z), user notes |
Model number, serial number, Mounting orientation (x, y, z), user notes |
|
| Export File Format |
CSV (comma-separated values) |
CSV (comma-separated values) |
|
| USB Port |
Loading Semi-Automating Test Routines Exporting Calibration Records in CSV Format |
Loading Semi-Automating Test Routines Exporting Calibration Records in CSV Format |
|
| Size - Width |
12 in |
30.5 cm |
|
| Size - Depth |
10 in |
28 cm |
|
| Size - Height |
8.5 in |
22 cm |
|
| Weight |
18 lb |
8.2 kg |
|
| Mounting Thread |
1/4-28 Female |
No Metric Equivalent |
|