high frequency cvld pressure sensor, 100 psi, 50 ua/psi,  acceleration compensated, 20 ft. integral coax cable, blunt cut, hydrotested

CVLD Pressure Sensor

Model 113M231

Retrieving Price...

(Pricing valid only in the U.S.)

Units in Stock Lead Times
More on the way Call for Lead Time
  • IP

High frequency CVLD pressure sensor, 100 psi, 50 uA/psi, acceleration compensated, 20 ft. integral coax cable, blunt cut, hydrotested

Ideal for hydro acoustic flow noise measurement
  • Measurement Range: 100 psi (689.5 kPa)
  • Sensitivity: (±15%)50 µA/psi (344.7 µA/kpa)
  • Low Frequency Response: (-5%)0.5 Hz
  • Resonant Frequency: >=200 kHz (>=200 kHz)
  • Weight: 0.51 oz (14.5 gm)

Specification Notes


[1] Zero-based, least-squares, straight line method.
[2] Typical.
[3] Bias mA = 2.1 x Excitation VDC - 22.8
[2] Typical.

English: SI:
Performance
Measurement Range 100 psi 689.5 kPa
Sensitivity (±15 %) 50 µA/psi 344.7 µA/kpa
Maximum Pressure 1000 psi 6895 kPa
Resonant Frequency ≥200 kHz ≥200 kHz
Rise Time ≤2.0 µ sec ≤2.0 µ sec
Low Frequency Response (-5 %) 0.5 Hz 0.5 Hz
Non-Linearity ≤1.0 % FS ≤1.0 % FS [1]
Environmental
Acceleration Sensitivity ≤0.005 psi/g ≤0.0035 kPa/(m/s²)
Temperature Range (Operating) -4 to +176 °F -20 to +80 °C
Temperature Coefficient of Sensitivity -0.03 %/°F -0.054 %/°C [2]
Maximum Vibration 2000 g pk 19614 m/s² pk
Maximum Shock 20000 g pk 196140 m/s² pk
Electrical
Output Polarity (Positive Pressure) Positive Positive
Discharge Time Constant ≥1.0 sec ≥1.0 sec
Excitation Voltage (±0.5 VDC) 15 VDC 15 VDC
Settling Time <7 sec <7 sec
Spectral Noise (1 kHz) 43 µpsi/√Hz 6192 µPa/√Hz [2]
Spectral Noise (10 kHz) 32 µpsi/√Hz 4608 µPa/√Hz [2]
Electrical Isolation ≥100000000 Ohm ≥100000000 Ohm
Bias Current 7 to 12 mA 7 to 12 mA [3]
Physical
Sensing Geometry Compression Compression
Sensing Element Quartz Quartz
Housing Material 17-4 Stainless Steel 17-4 Stainless Steel
Diaphragm 316L Stainless Steel 316L Stainless Steel
Coating Emralon Emralon
Sealing Welded Hermetic Welded Hermetic
Electrical Connections (center lead) Signal / Power Signal / Power
Electrical Connections (braid ) Ground Ground
Weight (without cable) 0.51 oz 14.5 gm [2]
Cable Termination Blunt cut Blunt cut
Cable Type Ø.12 santoprene jacket,coax,water blocked Ø.12 santoprene jacket,coax,water blocked
Cable Length 20 ft 6 m

* All specifications are at room temperature unless otherwise specified.