Model: 137B24B

$1,210.00 USD
(Pricing valid in US only)

Datasheet

Quartz, free-field, ICP® blast pressure pencil probe, 250 psi, 20 mV/psi, BNC connector

  • Measurement Range: 250 psi (1724 kPa)
  • Sensitivity: (±15%) 20 mV/psi (2.90 mV/kPa)
  • Resonant Frequency: >=400 kHz
  • Electrical Connector: BNC Jack
  • Weight: 12.4 oz (352 gm)

Add this product to my product comparison
Compare selected models ()

 English:SI: 
Performance
Measurement Range (for ±5V output)250 psi1724 kPa 
Measurement Range (for ±10V output)500 psi3447 kpsi[2] 
Sensitivity (±15 %)20 mV/psi2.90 mV/kPa 
Maximum Pressure5 kpsi34474 kPa 
Resolution.7 mpsi.005 kPa[3] 
Resonant Frequency≥400 kHz≥400 kHz 
Rise Time (Incident)≤6.5 µ sec≤6.5 µ sec 
Non-Linearity≤1.0 % FS≤1.0 % FS[1] 
Environmental
Temperature Range (Operating)-100 to +275 °F-73 to +135 °C 
Temperature Coefficient of Sensitivity≤0.05 %/°F≤0.090 %/°C 
Electrical
Discharge Time Constant (at room temp)≥0.2 sec≥0.2 sec 
Excitation Voltage20 to 30 VDC20 to 30 VDC 
Constant Current Excitation2 to 20 mA2 to 20 mA 
Output Impedance≤200 Ohm≤200 Ohm 
Output Bias Voltage8 to 14 VDC8 to 14 VDC 
Physical
Sensing GeometryCompressionCompression 
Sensing ElementQuartzQuartz 
Housing MaterialAluminumAluminum 
DiaphragmInvarInvar 
SealingEpoxyEpoxy 
Electrical ConnectorBNC JackBNC Jack 
Weight12.4 oz352 gm[3] 

All specifications are at room temperature unless otherwise specified.


CE

Product Notes
[1] Zero-based, least-squares, straight line method.
[2] For +10 volt output, minimum 26 VDC supply voltage required. Negative 10 volt output may be limited by output bias.
[3] Typical.
[4] See PCB Declaration of Conformance PS023 for details.
Accessories:
Supplied
PCS-6Calibration of Series 134, 137, and 138(1)
Optional Versions:
No optional versions
Product Documents
Additional Downloads