Log In

Log In

Need an Account? Start Here

 

Model: 134A24

Add to Cart for Quote

Tourmaline ICP® pressure bar, 1000 psi, 5 mV/psi, 0.2 microsecond rise time

  • Measurement Range: 1000 psi (6895 kPa)
  • Sensitivity: (±15%) 5 mV/psi (0.73 mV/kPa)
  • Low Frequency Response: (-5%) 0.25 Hz
  • Resonant Frequency: >=1500 kHz
  • Electrical Connector: 10-32 Coaxial Jack

Add this product to my product comparison
Compare selected models ()

 English:SI: 
Performance
Measurement Range (for ±5V output)1000 psi6895 kPa 
Useful Overrange (for ± 10V output)2000 psi13790 kPa[2] 
Sensitivity (±15 %)5 mV/psi0.73 mV/kPa[3] 
Maximum Pressure (static)15000 psi103425 kPa 
Resolution0.02 psi0.14 kPa 
Resonant Frequency≥1500 kHz≥1500 kHz 
Rise Time (Reflected)≤0.2 µ sec≤0.2 µ sec 
Low Frequency Response (-5 %)0.25 Hz0.25 Hz 
Non-Linearity≤2 % FS≤2 % FS[1] 
Environmental
Temperature Range (Operating)+32 to +120 °F0 to +49 °C 
Maximum Flash Temperature5000 °F2760 °C 
Electrical
Output Polarity (Positive Pressure)PositivePositive 
Discharge Time Constant (at room temp)≥1 sec≥1 sec 
Excitation Voltage20 to 30 VDC20 to 30 VDC 
Constant Current Excitation2 to 20 mA2 to 20 mA 
Output Impedance≤100 Ohm≤100 Ohm 
Output Bias Voltage8 to 14 VDC8 to 14 VDC 
Physical
Sensing ElementTourmalineTourmaline 
Housing MaterialStainless SteelStainless Steel 
DiaphragmEpoxyEpoxy 
SealingEpoxyEpoxy 
Electrical Connector10-32 Coaxial Jack10-32 Coaxial Jack 
Weight1.4 oz39 gm 

All specifications are at room temperature unless otherwise specified.


CE

Product Notes
[1] Zero-based, least-squares, straight line method.
[2] For +10 volt output, minimum 24 VDC supply voltage required. Negative 10 volt output may be limited by output bias.
[3] Calibrated as a system using (3 inch) type 003 cable and 402A series in-line amplifier.
[4] See PCB Declaration of Conformance PS023 for details.
Accessories:
Supplied
061A30Spanner Wrench, 2 Pin(1)
Optional Versions:
(M) Metric Mount
 ENGLISHSI
Product Documents
Additional Downloads